Adam J. Griebel, Jeremy E. Schaffer
Some implantable medical devices intended to treat maladies of the arteries of the brain, like flow diverters, require very thin wires, often 0.001” (25 µm) or less. Absorbable versions of these devices could offer several long-term advantages, but selecting a suitable material is challenging. Magnesium alloy wire at these sizes will likely degrade too quickly and be impossible to see with fluoroscopy. Zinc alloys are not elastic enough. Iron alloys, like our FeMnN alloy [1], may have an appropriate degradation rate but can still be prone to localized degradation, and would likely need radiopacity aids. Molybdenum (Mo) has recently been proposed as an absorbable metal [2] [3] and may offer a suitable corrosion rate and improved radiopacity, though much needs to be learned about its behavior in the body.
A new concept from Fort Wayne Metals R&D leverages our knowledge of absorbable metal alloy systems, our experience with producing DFT® wire composites, and our unique patent portfolio [4] [5]. We have produced several composite wires with FeMnN alloy shells and one or more Mo filaments embedded within. The galvanic relationship between FeMnN and Mo is such that the FeMnN will have its corrosion accelerated once the Mo is exposed, perhaps providing enough accelerant to overcome the passivating corrosion layer that plagues iron-based alloys. Based on prior work, FeMnN can provide enhanced elasticity and improved endothelial cell adhesion to promote reendothelialization [6, 7]. In turn, the Mo is galvanically protected from corrosion until the FeMnN is gone, providing a “programmable” staged degradation of the wire. Such wire might readily serve in absorbable flow diverters or other absorbable vascular devices.
Some of the prototype two-metal cross-sections produced are shown in Figure 1 below. Each of the cross-sections shows an outer shell layer of FeMnN surrounding one or an array of Mo core filaments. Trials to understand in vitro and in vivo degradation are underway, and work to characterize radiopacity is planned.
Click here to see previous highlights.
Disclaimer: Our monthly highlights are sneak peeks of what our R & D department is working on. This does not mean we have what is referenced above ready for manufacturing.
Adam J. Griebel, Jeremy E. Schaffer
Some implantable medical devices intended to treat maladies of the arteries of the brain, like flow diverters, require very thin wires, often 0.001” (25 µm) or less. Absorbable versions of these devices could offer several long-term advantages, but selecting a suitable material is challenging. Magnesium alloy wire at these sizes will likely degrade too quickly and be impossible to see with fluoroscopy. Zinc alloys are not elastic enough. Iron alloys, like our FeMnN alloy [1], may have an appropriate degradation rate but can still be prone to localized degradation, and would likely need radiopacity aids. Molybdenum (Mo) has recently been proposed as an absorbable metal [2] [3] and may offer a suitable corrosion rate and improved radiopacity, though much needs to be learned about its behavior in the body.
A new concept from Fort Wayne Metals R&D leverages our knowledge of absorbable metal alloy systems, our experience with producing DFT® wire composites, and our unique patent portfolio [4] [5]. We have produced several composite wires with FeMnN alloy shells and one or more Mo filaments embedded within. The galvanic relationship between FeMnN and Mo is such that the FeMnN will have its corrosion accelerated once the Mo is exposed, perhaps providing enough accelerant to overcome the passivating corrosion layer that plagues iron-based alloys. Based on prior work, FeMnN can provide enhanced elasticity and improved endothelial cell adhesion to promote reendothelialization [6, 7]. In turn, the Mo is galvanically protected from corrosion until the FeMnN is gone, providing a “programmable” staged degradation of the wire. Such wire might readily serve in absorbable flow diverters or other absorbable vascular devices.
Some of the prototype two-metal cross-sections produced are shown in Figure 1 below. Each of the cross-sections shows an outer shell layer of FeMnN surrounding one or an array of Mo core filaments. Trials to understand in vitro and in vivo degradation are underway, and work to characterize radiopacity is planned.
Click here to see previous highlights.
Disclaimer: Our monthly highlights are sneak peeks of what our R & D department is working on. This does not mean we have what is referenced above ready for manufacturing.